Overview

MDS20B-24 is the 24V output model of this series: rated power 20W, typical efficiency 89%.

MDS20 series is 20W isolated and regulated module type medical grade DC-DC converter with 2"x1" package. It features international standard pins, a high efficiency up to 90%, wide working temperature range -40~+90℃, 4KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~18V, 18~36V and 36~75V 2:1 wide input range, and various output voltage 5V/12V/15V/24V, which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 2"x1" package with international standard pinout
  • 2:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • No minimum load required
  • Protections: Short circuit (Continuous) / Overload / Input under voltage
  • 4000VAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Specifications

MDS20B-24Specifications from the official datasheet (2024-04-30 edition)

INPUT

INPUT VOLTAGE (RANGE)
Normal 24V (18 ~ 36V)
INPUT CURRENT NO LOAD
7mA
INPUT CURRENT FULL LOAD
1000mA
VOLTAGE RANGE
18~36Vdc
SURGE VOLTAGE (100ms max.)
12Vin models : 25Vdc, 24Vin models : 50Vdc, 48Vin models : 100Vdc
FILTER
Pi type
PROTECTION
Fuse recommended. 12Vin models: 5A delay time Type, 24Vin models: 3A delay time Type, 48Vin models: 2A delay time Type

OUTPUT

OUTPUT VOLTAGE
24V
OUTPUT CURRENT
833mA
EFFICIENCY (Typ.)
89%
CAPACITOR LOAD (MAX.)
300μF
VOLTAGE ACCURACY
±1%
RATED POWER
20W
RIPPLE & NOISENote.2
60mVp-p
LINE REGULATIONNote.3
±0.5%
LOAD REGULATIONNote.4
±0.5%
SWITCHING FREQUENCY (Typ.)
12Vin/24Vin:250KHz, 48Vin:300KHz

PROTECTION

SHORT CIRCUIT
Protection type : Continuous, automatic recovery
OVERLOAD
110 ~ 180% rated output power Protection type : Recovers automatically after fault condition is removed
UNDER VOLTAGE LOCKOUT
12Vin: 7.5Vdc, 24Vin: 15Vdc, 48Vin: 33Vdc

ENVIRONMENT

COOLING
Free-air convection
WORKING TEMP.
-40 ~ +90℃ (Refer to "Derating Curve")
CASE TEMPERATURE
+105℃ max.
WORKING HUMIDITY
20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY
-55 ~ +125℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT
0.03% / ℃ (0 ~ 70℃)
SOLDERING TEMPERATURE
1.5mm from case of 1 ~ 3sec./260℃ max.
VIBRATION
10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes

SAFETY & EMC ( Note.6)

SAFETY STANDARDS
IEC60601-1(LVD) 3rd edition, EAC TP TC 020/2011(EAC TP TC 004 for 48Vin type only) approved
WITHSTAND VOLTAGE
I/P-O/P 4KVAC
ISOLATION RESISTANCE
I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
ISOLATION CAPACITANCE (Typ.)
100pF
EMC EMISSION
ParameterStandardTest Level
ConductedBS EN/EN55011Class A without external components
RadiatedBS EN/EN55011Class A without external components
EMC IMMUNITY
ParameterStandardTest Level
ESDBS EN/EN61000-4-2Level 1, ±15KV air, ±8KV contact
Radiated SusceptibilityBS EN/EN61000-4-3Level 1, 10V/m
EFT/BurstsBS EN/EN61000-4-4Level 1, ±2KV
SurgeBS EN/EN61000-4-5Level 1, ±1KV Line-Line, ±2KV Line-GND
ConductedBS EN/EN61000-4-6Level 1, 10Vrms(e.m.f.)

OTHERS

MTBF
1016Khrs MIL-HDBK-217F(25℃)
DIMENSION (L*W*H)
50.8*25.4*12mm (2*1*0.47 inch)
CASE MATERIAL
UL94V-0 plastic case
PACKING
30g; 18pcs/per tube, 432pcs/24 tube max./carton
Test conditions and notes6
  1. 1.All parameters are specified at normal input(A:12Vdc, B:24Vdc, C:48Vdc), rated load, 25℃ 70% RH ambient.
  2. 2.Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μf & 47μf capacitor.
  3. 3.Line regulation is measured from low line to high line at rated load.
  4. 4.Load regulation is measured from 0% to 100% rated load.
  5. 5.2xMOPP base on a working voltage of 250Vac between Primary to Secondary.
  6. 6.The final equipment must be re-confirm that it still meet EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”(as available on http://www.meanwell.com)

Technical drawings

MDS20B-24 Pin Configuration
Pin Configuration
MDS20B-24 Derating Curve
Derating Curve
MDS20B-24 Packing
Packing

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment

Downloads

FAQ

Can I use the MDS20B-24 for 24V industrial controls on an 18-36V input system?

Yes, this converter is designed for 24V constant voltage output from an 18-36V input range. Its total output power is 20W, so it is best matched to lower-power 24V control, monitoring, or automation loads within that rating.

Is the MDS20B-24 appropriate for 24V sensors or interface modules inside automation equipment?

It can be used for 24V electronics that need constant voltage and fit within the converter's 20W capacity. The compact 50.8 × 25.4 × 12.0 mm housing also makes it suitable for dense control assemblies.

Can this converter power 24V embedded or industrial communication devices?

It is intended for 24V loads when the source is within 18-36V and the total demand does not exceed 20W. That makes it a candidate for small 24V embedded or industrial communication hardware rather than higher-power 24V equipment.

What are the differences between the MDS20B-24 and the other MDS20B versions?

All listed MDS20B versions are 20W converters with the same 18-36V input range. The outputs are different: MDS20B-05 is 5V at 4A, MDS20B-12 is 12V at 1.67A, MDS20B-15 is 15V at 1.33A, and MDS20B-24 is 24V at 833mA.

Can the MDS20B-24 be used in outdoor or temperature-extreme applications?

The specified operating temperature range is -40°C to +90°C, with output derating at higher ambient temperatures per the datasheet derating curve. No IP rating appears in the provided specs, so these specs do not indicate exposed outdoor, wet, or washdown suitability.

Payment & Security

Payment methods

  • American Express
  • Apple Pay
  • Bizum
  • Diners Club
  • Discover
  • Google Pay
  • Mastercard
  • PayPal
  • Shop Pay
  • Visa

Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.

{ "id":8658947244352, "variants" :[{ "context":"default", "variant_id": 46633509781824, "sku":"MDS20B-24", "inventory_qty" : 0, "customer" : true, "custom_html" :"\u003cdiv class=\"price\"\u003e\n \u003cdiv class=\"price__default\"\u003e\n \u003cstrong class=\"price__current\"\u003e$40.29\u003c\/strong\u003e\n \u003cs class=\"price__was\"\u003e\u003c\/s\u003e\n \u003c\/div\u003e\n\n \u003cdiv class=\"unit-price relative\" hidden\u003e\u003cspan class=\"visually-hidden\"\u003eUnit price\u003c\/span\u003e\u003cspan class=\"unit-price__price\"\u003e\u003c\/span\u003e\u003cspan class=\"unit-price__separator\"\u003e \/ \u003c\/span\u003e\u003cspan class=\"unit-price__unit\"\u003e\u003c\/span\u003e\u003c\/div\u003e\n\n \u003cdiv class=\"price__no-variant\" hidden\u003e\n \u003cstrong class=\"price__current\"\u003eUnavailable\u003c\/strong\u003e\n \u003c\/div\u003e\n\n \u003c\/div\u003e\n", "volume_html" :"\u003cvolume-pricing class=\"parent-display inside_product_form\" id=\"volume-pricing-main\" \u003e\u003c\/volume-pricing\u003e" } ] }
Mean Well MDS20B-24 DC-DC Converter - 20W - 18~36V in 24V out 833mA
Mean Well

Mean Well MDS20B-24 DC-DC Converter - 20W - 18~36V in 24V out 833mA

$40.29 USD

MDS20B-24 is the 24V output model of this series: rated power 20W, typical efficiency 89%.

MDS20 series is 20W isolated and regulated module type medical grade DC-DC converter with 2"x1" package. It features international standard pins, a high efficiency up to 90%, wide working temperature range -40~+90℃, 4KVAC I/P-O/P high isolation voltage, compliance with IEC60601-1 medical standard, continuous-mode short circuit protection, etc. The models account for different input voltage 9~18V, 18~36V and 36~75V 2:1 wide input range, and various output voltage 5V/12V/15V/24V, which are suitable for medical systems, ultra low leakage current.

Features

  • DIP 2"x1" package with international standard pinout
  • 2:1 wide input range
  • Low patient leakage current <5μA
  • Wide operating temperature range -40 ~ +90℃
  • Medical safety approved (2xMOPP) according to IEC60601-1
  • No minimum load required
  • Protections: Short circuit (Continuous) / Overload / Input under voltage
  • 4000VAC high I/O isolation (Reinforced isolation)
  • 3 years warranty

Typical applications

  • Medical devices
  • Medical oxygen monitor
  • CT scanning
  • Medical carts
  • Oral care equipment
View product