Over 15 years experience of providing support of MEAN WELL Power Supplies
Overview
GSM40A12-P1J is the 12V output model of this series: rated current 3.34A, typical efficiency 88%.
GSM40A is a highly reliable, 40W single-output green medical adaptor series. This product is a class I power unit(with FG), equipped with a standard IEC320-C14 AC inlet and adopting the input range from 80VAC to 264VAC. The entire series supplies different models with output voltages between 5VDC and 48VDC that can satisfy the demands for various types of medical electrical devices. The circuitry design meets the international medical standards (2*MOPP), having an ultra low leakage current (<100μA), fitting the medical devices in direct electrical contact with the patients. With the efficiency up to 91% and the extremely low no-load power consumption below 0.15W, GSM40A is compliant with USA EISA 2007/DoE ,Canada NRCan,Australia and New Zealand MEPS, EU ErP,and meet Code of Conduct (CoC) Version 5.The supreme feature allows the adaptor to save the energy when it is either under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic case.GSM40A is certified for the international medical safety regulations.
Features
- 3 pole AC inlet IEC320-C14, ClassⅠpower unit
- Medical safety approved (2 x MOPP) according to ANSI/AAMI ES60601-1 and IEC/BS EN/EN60601-1
- Extremely low leakage current
- No load power consumption<0.15W
- Energy efficiency level VI and meet CoC Version 5 (Except 5~9V for Level V)
- -30~+70℃ wide range working temperature
- Protections: Short circuit / Overload / Over voltage
- LED indicator for power on
- Lifetime > 90 K hours
- Various DC plug quick adapter accessory available (Plug kit sold sperately,please refer to : https://www.meanwell.com/upload/pdf/DC_plug.pdf)
- 3 years warranty
Specifications
OUTPUT
- SAFETY MODEL NO.
- GSM40A12
- DC VOLTAGENote.2
- 12V
- RATED CURRENT
- 3.34A
- CURRENT RANGE
- 0.1 ~ 3.34A
- RATED POWER (max.)
- 40W
- RIPPLE & NOISE (max.)Note.3
- 100mVp-p
- VOLTAGE TOLERANCENote.4
- ±3.0%
- LINE REGULATIONNote.5
- ±1.0%
- LOAD REGULATION
- ±3.0%
- SETUP, RISE TIMENote.6
- 1000ms, 30ms / 230VAC 1500ms, 30ms / 115VAC at full load
- HOLD UP TIME (Typ.)
- 50ms / 230VAC 24ms / 115VAC at full load
INPUT
- VOLTAGE RANGENote.7
- 80 ~ 264VAC
- FREQUENCY RANGE
- 47 ~ 63Hz
- EFFICIENCY (Typ.)
- 88%
- AC CURRENT (Typ.)
- 1A / 115VAC 0.5A / 230VAC
- INRUSH CURRENT (Typ.)
- Cold start 30A/115VAC 60A / 230VAC
- LEAKAGE CURRENT(max.)
- Earth leakage current < 100μA/264VAC , Touch current < 100μA/264VAC
PROTECTION
- OVERLOAD
- 105 ~ 160% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed
- OVER VOLTAGE
- 12.6 ~ 16.2V Protection type : Shut down o/p voltage, re-power on to recover
ENVIRONMENT
- WORKING TEMP.
- -30 ~ +70℃ (Refer to "Derating Curve")
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -40 ~ +85℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- ±0.03% / ℃ (0~50℃ )
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
- OPERATING ALTITUDENote.8
- 3000 meters
SAFETY & EMC (Note 9)
- SAFETY STANDARDS
- IEC 60601-1:2005+A1+A2, TUV BS EN/ EN 60601-1:2006+A1+A12+A2, ANSI/AAMI ES60601-1:2005+A2,CAN/CSA C22.2 No. 60601-1:2014+A2, EAC TP TC 004 approved
- ISOLATION LEVEL
- Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP
- WITHSTAND VOLTAGE
- I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:SHORT
- ISOLATION RESISTANCE
- I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25℃ / 70% RH
- EMC EMISSION
-
Parameter Standard Test Level Conducted emission BS EN/EN55011 (CISPR11), FCC PART 15 / CISPR22 Class B Radiated emission BS EN/EN55011 (CISPR11), FCC PART 15 / CISPR22 Class B Harmonic current BS EN/EN61000-3-2 Class A Voltage flicker BS EN/EN61000-3-3 - EMC IMMUNITY
-
BS EN/EN60601-1-2, BS EN/EN61204-3
Parameter Standard Test Level ESD BS EN/EN61000-4-2 Level 4, 15KV air; Level 4, 8KV contact RF field susceptibility BS EN/EN61000-4-3 Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz ) EFT bursts BS EN/EN61000-4-4 Level 3, 2KV Surge susceptibility BS EN/EN61000-4-5 Level 3, 1KV/Line-Line , 2KV/Line-FG Conducted susceptibility BS EN/EN61000-4-6 Level 3, 10V Magnetic field immunity BS EN/EN61000-4-8 Level 4, 30A/m Voltage dip, interruption BS EN/EN61000-4-11 100% dip 1 periods, 30% dip 25 periods, 100% interruptions 250 periods
OTHERS
- MTBF
- 3613.2K hrs min. Telcordia SR-332 (Bellcore) ; 740.7K hrs min. MIL-HDBK-217F (25℃)
- DIMENSION
- 125*50*31.5mm (L*W*H)
- PACKING
- 0.25Kg; 40pcs/ 12Kg/1.04CUFT
CONNECTOR
- PLUG
- See page 4~5 ; Other type available by customer requested
- CABLE
- See page 4~5 ; Other type available by customer requested
Test conditions and notes9
- 1. All parameters are specified at 230VAC input, rated load, 25℃ 70% RH ambient.
- 2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
- 3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μF & 47μF capacitor.
- 4. Tolerance: includes set up tolerance, line regulation, load regulation.
- 5. Line regulation is measured from low line to high line at rated load.
- 6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
- 7. Derating may be needed under low input voltages. Pleas check the derating curve for more details.
- 8. The ambient temperature derating of 3.5℃/1000m with fanless models and of 5℃/1000m with fan models for operating altitude higher than 2000m(6500ft).
- 9. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )
Technical drawings
Typical applications
- Mobile clinical workstation
- Oral irrigator
- Portable hemodialysis machine
- Breath Machine
- Medical computer monitor
Downloads
Datasheet1
Certificates7
FAQ
What output does the Mean Well GSM40A12-P1J provide?
The GSM40A12-P1J is a 40W AC-DC medical adaptor with a regulated 12V DC output. It is rated for up to 3.34A output current. This makes it suitable for equipment that requires a 12V supply within that power range.
Is the GSM40A12-P1J approved for medical applications?
Yes. This model is medical safety approved and specified as providing 2 x MOPP isolation between primary and secondary. It also lists compliance with medical safety standards including IEC 60601-1, ANSI/AAMI ES60601-1 and EN 60601-1.
What AC input range does this power supply support?
The GSM40A12-P1J has a universal input range of 80 to 264VAC at 47 to 63Hz. That wide range helps it fit international mains environments. It uses an IEC320-C14 AC inlet.
What DC plug type comes standard on the GSM40A12-P1J?
The standard output plug for this model is the P1J type. It is specified as 2.1 mm inner diameter by 5.5 mm outer diameter by 11 mm length, center positive, tuning fork type. Buyers should match that plug format to their device requirements.
What protection features are included in this model?
This power supply includes overload and over voltage protection. The overload protection is specified at 105% to 160% of rated output power and uses hiccup mode with auto-recovery, while over voltage protection (12.6 to 16.2V) shuts down the output until the unit is re-powered.
What operating environment is the GSM40A12-P1J designed for?
The specified working temperature range is -30°C to +70°C, with reference to the output derating curve. Working humidity is 20% to 90% RH non-condensing, and the unit includes a power LED indicator. It is rated for operating altitudes up to 3000 meters.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Please log in to write a review.
Write a review
Reviews are checked before publishing. By submitting you confirm this is your genuine experience with this product.
Submitting your review…
It's published and now helps other buyers.
We're checking it and it'll appear shortly.
Please try again.