Over 15 years experience of providing support of MEAN WELL Power Supplies
Overview
GSM36U48-P1J is the 48V output model of this series: rated current 0.75A, typical efficiency 88%. Type P1J: Plug for standard model, 2.1φ x 5.5φ x11mm, c+, tuning fork type.
GSM36U is a highly reliable, 36W wall-mounted style single-output green medical adaptor series. This product is equipped with a 2-pin (no FG) standard USA AC power plug, adopting the input range from 80VAC to 264VAC. The entire series supplies different output voltages between 5VDC and 48VDC that can satisfy the demands for various kinds of miniature medical devices. The circuitry design meets the international medical standards (2 × MOPP), having an ultra low leakage current (<50 μA), fitting the medical devices in direct electrical contact with the patients. With the efficiency up to 88% and the extreme low no-load power consumption below 0.1W, GSM36U is compliant with USA EISA 2007/DoE and Canada NRCan; the superme feature allows the adaptor to save the energy when it is either under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic case, providing the double insulation that effectively prevents electrical shock. GSM36U is approved with the international medical safety certificates.
Features
- 2 pole USA plug, Class Ⅱ power unit
- Medical safety approved (2 × MOPP) according to ANSI/AAMI ES60601-1/-1-11
- Extremely low leakage current
- No load power consumption<0.1W
- Energy efficiency Level Ⅵ ( except 5~9V for Level Ⅴ)
- -25~+60℃ wide range working temperature
- Protections: Short circuit / Overload / Over voltage
- LED indicator for power on
- Various DC plug quick adapter accessory available (Plug kit sold sperately, please refer to : https://www.meanwell.com/upload/pdf/DC_plug.pdf)
- 3 years warranty
Specifications
OUTPUT
- SAFETY MODEL NO.
- GSM36U48
- DC VOLTAGENote.2
- 48V
- RATED CURRENT
- 0.75A
- CURRENT RANGE
- 0 ~ 0.75A
- RATED POWER (max.)
- 36W
- RIPPLE & NOISE (max.)Note.3
- 240mVp-p
- VOLTAGE TOLERANCENote.4
- ±2.0%
- LINE REGULATIONNote.5
- ±1.0%
- LOAD REGULATION
- ±2.0%
- SETUP, RISE TIMENote.6
- 500ms, 30ms / 230VAC 1000ms, 30ms / 115VAC at full load
- HOLD UP TIME (Typ.)
- 16ms / 230VAC 10ms / 115VAC at full load
INPUT
- VOLTAGE RANGENote.7
- 80 ~ 264VAC 113 ~ 370VDC
- FREQUENCY RANGE
- 47 ~ 63Hz
- EFFICIENCY (Typ.)
- 88%
- AC CURRENT (Typ.)
- 0.9A / 115VAC 0.45A / 230VAC
- INRUSH CURRENT (Typ.)
- 55A / 230VAC 30A / 115VAC
- LEAKAGE CURRENT(max.)
- Touch current < 50μA/264VAC
PROTECTION
- OVERLOAD
- 105 ~ 170% rated output power Protection type : Hiccup mode, recovers automatically after fault condition is removed
- OVER VOLTAGE
- 50.4 ~ 64.8V Protection type : Shut down o/p voltage, re-power on to recover
ENVIRONMENT
- WORKING TEMP.
- -25 ~ +60℃ (Refer to "Derating Curve")
- WORKING HUMIDITY
- 20% ~ 90% RH non-condensing
- STORAGE TEMP., HUMIDITY
- -40 ~ +85℃, 10 ~ 95% RH non-condensing
- TEMP. COEFFICIENT
- ±0.03% / ℃ (0~40℃)
- VIBRATION
- 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes
SAFETY & EMC (Note. 8)
- SAFETY STANDARDS
- IEC 60601-1:2005+A1+A2; IEC 60601-1-11:2015+A1, ANSI/AAMI ES60601-1:2005+A2; ANSI/AAMI HA60601-1-11+A1 CAN/CSA C22.2 No. 60601-1:2014+A2; CSA C22.2 NO. 60601-1-11:2015+A1, EAC TP TC 004 approved
- ISOLATION LEVEL
- Primary-Secondary: 2xMOPP
- WITHSTAND VOLTAGE
- I/P-O/P:4KVAC
- ISOLATION RESISTANCE
- I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH
- EMC EMISSION
-
Parameter Standard Test Level Conducted emission FCC PART 15 / CISPR22 Class B Radiated emission FCC PART 15 / CISPR22 Class B
OTHERS
- MTBF
- 3777.8K hrs min. Telcordia SR-332 (Bellcore) ; 651.4K hrs min. MIL-HDBK-217F (25℃)
- DIMENSION
- 79*54*33mm (L*W*H)
- PACKING
- 215g ; 60pcs / 13.9Kg / CARTON
CONNECTOR
- PLUG
- See page 4~5 ; Other type available by customer requested
- CABLE
- See page 4~5 ; Other type available by customer requested
MODEL ENCODING
- P1J (this model)
- Plug for standard model, 2.1φ x 5.5φ x11mm, c+, tuning fork type
- Blank
- Other options available by customer requested (see Page 4~5)
Test conditions and notes8
- 1. All parameters are specified at 115VAC input, rated load, 25℃ 70% RH ambient.
- 2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.
- 3. Ripple & noise are measured at 20MHz by using a 12" twisted pair terminated with a 0.1μF & 47μF capacitor.
- 4. Tolerance: includes set up tolerance, line regulation, load regulation.
- 5. Line regulation is measured from low line to high line at rated load.
- 6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.
- 7. Derating may be needed under low input voltage. Please check the derating curve for more details.
- 8. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.” (as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )
Technical drawings
Typical applications
- Blood glucose meter
- Blood pressure meter
- Nebulizer
- Inhaler
- Portable medical device
Downloads
Datasheet1
Certificates5
Test reports1
FAQ
Can I use the GSM36U48-P1J for 48V industrial control hardware?
It can fit 48V DC control or embedded hardware that requires constant voltage and draws no more than 0.75A. This model is rated for 36W total output, so the connected 48V load needs to stay within that power and current limit.
Is the GSM36U48-P1J suitable for small 48V telecom or network equipment?
It can be suitable for 48V equipment that uses a constant voltage input and stays within 36W and 0.75A. The wide 80-264VAC input also helps when the AC source varies across regions or facilities.
Can this 48V power supply run embedded systems or test fixtures that need stable DC output?
Yes, if the embedded system or fixture is designed for 48V constant voltage power and its draw does not exceed 0.75A. The compact 79.0 x 54.0 x 33.0 mm size can also help where space is limited.
Is the GSM36U48-P1J a good fit for low-power 48V automation accessories?
It is appropriate for low-power 48V automation accessories when the total load stays within 36W. For selection, the key checks are that the device requires 48V constant voltage and that its current demand is at or below 0.75A.
Can the GSM36U48-P1J be used outdoors or in harsh plant conditions?
The provided specs list an operating temperature range of -25°C to +60°C, but they do not list an IP rating. That means the temperature capability is defined, but outdoor, washdown, or dust-tight suitability is not established by the stated specs.
Payment & Security
Payment methods
Your payment information is processed securely. We do not store credit card details nor have access to your credit card information.
Other GSM36 Models
1 other model| Model | Power | Voltage | Current | |
|---|---|---|---|---|
| Mean Well GSM36U12-P1J Power Supply 36W 12V 3A | 36W | 12V | 3A | View |
Please log in to write a review.
Write a review
Reviews are checked before publishing. By submitting you confirm this is your genuine experience with this product.
Submitting your review…
It's published and now helps other buyers.
We're checking it and it'll appear shortly.